Dong-Sik Chae
Research Interests
Explore related searches
Contact this professor
Get a free first email built on one of their papers, ready for you to edit and send yourself.
Articles (1)
Background: Fractures of the femur require significant radiation exposure during operations using fluoroscopy (C-arm), posing a high risk of radiation exposure to the medical staff and patients. To address this concern, in this study, we investigated the efficacy of using two fluoroscopy machines simultaneously. Methods: We categorized 30 patients with femoral neck fracture (FNF) into single and double C-arm groups. The operation and radiation exposure times during a closed reduction and internal fixation operation were investigated to evaluate whether the operation and radiation exposure times were effectively audited when the operation was performed using a double C-arm. Results: The total operation times were 93.21 ± 20.70 min and 66.69 ± 13.97 min for the single and double C-arm groups, respectively. Additionally, the total radiation times were 100.43 ± 24.59 s and 83.06 ± 19.53 s for the single and double C-arm groups, respectively. Operation and radiation exposure times in the two groups showed statistically significant differences (p < 0.05). Conclusion: The use of double C-arm in FNF can reduce operation and radiation exposure times. Hence, using the double C-arm in surgical treatment could reduce the risk of radiation exposure to medical staff and patients.
Year:
2024
Collaborators (2)
Seung-Yong Sung
-
HYUN HEE LEE
Clinical Assistant Professor
Catholic Kwandong University International Saint Mary's Hospital
Make your next move.
Prepare your application
Review drafts before you send them.